Designation, Purpose, Use of the Equipment
The device is intended to secure a safe and simple manner of manipulation with radiopharmaceuticals during preparation of the dose and administering of radioactive solution of radiopharmaceuticals to the patient. Its purpose is to limit the risk of irradiation and bacterial contamination of employees when handling radioactive solutions.
It facilitates maintaining aseptic conditions when manipulating with the syinge and needle during administration procedure.

Main components of the device
The body of the shielding is made of tungsten. The stainless-steel screw with grooved head provides the fixation of the syringe inside the shielding. The window with shielding glass enables to watch the piston of the syringe.
Main dimensions PT358
- Inside diameter 15,2 mm
- Lenght 65 mm
- Outside diameter 34 mm
Main dimensions PT359
- Inside diameter 12,4 mm
- Lenght 65 mm
- Outside diameter 32,6 mm
Basic weight PT358
- 850 g
Basic weight PT359
- 825 g
Related devices (are not a part of this device)
- Automatic batching and measuring device PT317
- Administering stand PT353
- Simple stand for connecting a needle or stopper PT326
- Petomobile PT151
Accessories Supplied with the Device
- none
----------------------------------------------------
Safe and easy individual dose preparation
- Prepare an individual dose by just a couple of clicks! The dose will be ready within 1 minute with the accuracy better than 3 %.
- The built-in measuring system measures the activity in the syringe directly during the dose preparation process and displays it on the control PC screen.
- The activity remaining in mother vial is calculated and displayed similarly.
- PET doses can be prepared into 3 ml, 5 ml and 10 ml syringes.

Shielding parameters
- The vial shielded by eqv. of 68mm Pb - Combination of W and Pb shield
- Syringes shields - range from 8 to 11mm W

Tungsten Container PT318R2
specially designed tungsten container fits the packaging set Type A PT314R2 for RF transportation. Using this solution eliminates the unshielded manipulation with an activity containing vial.

Highly cost effective
- Combination of W and Pb shielding materials
- Vial in a slanted position - no waste of RF (patented solution)
- Only the standard consumables are used (BBraun or others)
- No need to use any special disposable kits
- Quick and easy wastes handling (only two needles and a vial)
- Reasonable level of automation
- Minimum time for the first dose set up needed, max. 3 min.
- Automatic daily QC procedure with built in check emitter
Intuitive and user-friendly software
Software continuously displays the remaining activity in the vial (calculated value). Automatic linearity and stability tests - check emitter incorporated in the device body. The software will recommend a value of individual dose according to the patient’s weight or BMI based on preset specifications. Selection of automatic, semiautomatic or manual modes.

Consumables for one vial
- 1x BBraun Spinocan 0,90 x 88 mm; G20
- 1x BBraun Sterican 0,80 x 120 mm; G21

Technical specifications
- Dose accuracy ..... ±3%
- Dose range ..... 35 -1800 MBq
- Max. vial volume ..... 25 ml
- Max. activity in the vial ..... 20 GBq
- Max. concentration ..... 1 500 MBq/ml
- Radiopharmaceuticals ..... All PET
Dimensions
- Height ..... 450 mm
- Width ..... 380 mm
- Length ..... 299 mm